Product Overview
The RKI Instruments 43-4170RK is a fast-acting 6 A, 250 V cartridge fuse, 1/4" x 1-1/4" (about 6.3 x 32 mm). RKI’s Digester Gas Monitor manual lists it as the AC fuse.
| Specification | Value |
|---|---|
| Rating | 6 A, 250 V |
| Speed | Fast-acting |
| Size | 1/4" x 1-1/4" |
| Application | RKI Digester Gas Monitor AC fuses F1 and F2 |
Where It Goes
The Digester Gas Monitor uses two AC fuses (F1, F2) and one DC fuse (F3). They sit in vertical fuse holders with quarter-turn covers on the right side of the main board, between the power switch and the AC input terminal strip.
- F1, F2 (AC): 6 A, 250 V, 1/4" x 1-1/4", fast-acting (43-4170RK)
- F3 (DC): 10 A, 250 V, 1/4" x 1-1/4", fast-acting (43-4179RK)
Before You Replace a Fuse
- A blown fuse usually means a fault. Find the cause before installing a new one.
- Check the rating marked on the fuse or circuit board, and replace with the same rating and type.
- RKI’s 2026 price list says the Digester Gas Monitor was redesigned in November 2024 and is now powered by 24 VDC only. The fuse layout above comes from the January 2024 manual (Rev. S), which covers AC and 24 VDC power. Confirm the fuses in a newer unit against its manual.
References
- RKI Instruments 2026 Price List (02/17/2026)
- RKI Digester Gas Monitor Operator’s Manual 71-0266RK (Rev. S, released 1/22/24), Chapter 2 (Fuses) and Table 19 parts list
(P/N: 43-4170RK)