RKI Instruments Hydrogen Sulfide (H2S) Sensor, 0-200 ppm / 1,000 ppm, Digester Gas Monitor - 65-2039RK

Price: $577.00
SKU:
65-2039RK
Model:
Digester Gas Monitor
Adding to cart… The item has been added

Product Overview

The RKI Instruments 65-2039RK is the plug-in hydrogen sulfide (H2S) replacement sensor for the RKI Digester Gas Monitor (72-2120A series). RKI’s price list describes it as “Sensor, Hydrogen Sulfide (H2S) 0-200 ppm / 1,000 ppm.”

The monitor blends the sample with fresh dilution air ahead of the H2S sensor. That is how the H2S channel is set up for full scales of 0–500, 0–1,000, 0–3,000 or 0–5,000 ppm. The dilution flow settings depend on the range, so keep the flowmeter settings that match your unit.

Compatible Instrument

  • RKI Digester Gas Monitor, versions with an H2S channel (for example 72-2120A-401, -301, -303, -304, -202, -204 and -103)

Digester H2S Channel Settings

H2S rangeResolutionH2S dilution sample flowmeterH2S dilution air
0–500 ppm1 ppm61 mm (0.4 SCFH)1.2 SCFH
0–1,000 ppm5 ppm50 mm (0.3 SCFH)1.0 SCFH
0–3,000 ppm10 ppm44 mm (0.25 SCFH)1.2 SCFH
0–5,000 ppm20 ppm36 mm (0.2 SCFH)1.4 SCFH

The Flow to Sensors flowmeter should always read 3.0 SCFH.

Installation

  1. Turn off the monitor and its incoming power, open the door, and turn the flow regulator knob fully counterclockwise.
  2. Remove the two screws holding the H2S preamp board (front left of the flow block). Keep the washers, and make sure the gasket stays in the sensor cavity.
  3. Pull the old sensor off the preamp board.
  4. Remove the shipping wire jumper from the new sensor’s pins. RKI warns the sensor will not work properly if the jumper is left on.
  5. Plug the new sensor into the preamp, reinstall the board, restore power, and set the sensor flow to 3 SCFH.
  6. Calibrate the new sensor.

Calibration

  • Calibrate the H2S channel separately from the methane, oxygen and CO2 channels.
  • Use H2S span gas between 10% and 100% of the channel’s full scale, with a demand-flow regulator and non-absorbent tubing.
  • RKI’s Digester parts list includes 81-0145RK-02 (500 ppm H2S, 58 L) and 81-0146RK-02 (200 ppm H2S, 58 L), plus the 81-1054RK demand-flow regulator.
  • RKI’s typical calibration interval for the Digester Gas Monitor is quarterly.
  • Safety: RKI warns that H2S calibration gas at these concentrations is dangerous if inhaled. Route the monitor’s exhaust line to a safe area.

References