PS200
Overview:
A robust and accurate portable gas detector, the Personal Surveyor (PS200 Series) provides unrivalled protection in confined space applications with audible and visual alarms in the event of exposure to flammable or toxic gases. Detecting and displaying up to 4 gases simultaneously, PS200 is suitable for a host of applications in a variety of industries. The PS200 can be configured to detect a combination of Methane, Oxygen, Carbon Monoxide, Hydrogen Sulphide, as well as other flammable gases. Pre-entry checking can be carried out with the optional internal sampling pump, and diffusion operation ensuring maximum battery life in confined spaces.
Key Features:
- Simple 1-button operation
- Measures up to 4 gases:
- CH4 – Methane (LEL)
- O2 – Oxygen
- CO – Carbon Monoxide
- H2S – Hydrogen Sulphide
- Other flammable gases, available on request
- Audible, visual and vibrating alarms
- Internal pump (optional)
- Lightweight yet robust construction
- Sensor integrity checking
- Ease of maintenance
Applications:
- Personal safety in confined space – CH4/O2/CO/H2S hazards
- General safety in confined space
- Shipping – general
- Shipping – fumigant
- Breweries
- Plant shutdown
- Permit to work systems
- Pressurised tunnelling
- Fleet auxiliary
PS200 calibration guide
MANUAL CALIBRATION allows you to calibrate using any cal gas values. Apply the gas and then adjust the values of the sensors to match. To manual calibrate:
- Turn on
- Wait for countdown screen to start
- Immediately press left right left right
- Wait for countdown to complete
- Go to calibrate
- Code 333
- Power to zero
- Go to user set
- Turn the knob on the 0.5 lpm fixed flow regulator to calibrate
- Use left button to alter when stabilized, use right button to accept
Docking-station gas: The PS200 Auto Bump & Calibration (ABC) Station is configured by default for GMI Combi Test Gas 99146: 2.5% vol methane (50% LEL), 500 ppm CO, 50 ppm H2S, 18% vol O2, balance nitrogen. Our equivalent is Blend 0480. The manual-calibration values in the table below are separate from the dock setup. Whatever cylinder you use, the values set in the instrument or station must match the cylinder label.
| Sensor Type | Calibration Gas Value | Low Alarm | High Alarm | TWA Alarm | STEL Alarm | Part Number |
|---|---|---|---|---|---|---|
| H2S - Hydrogen Sulfide | 25ppm | 10ppm | 20ppm | 10ppm | 15ppm | 630-64067DD |
| CO - Carbon Monoxide | 100ppm | 35ppm | 70ppm | 35ppm | 200ppm | 630-64067DD |
| LEL - Methane Curve | 50% LEL | 10% LEL | 20% LEL | N/A | N/A | 630-66825 |
| O2 - Oxygen | 18% VOL | 19.5% VOL | 23.5% VOL | N/A | N/A | 630-64066 |