G888

GfG Microtector III G888C General Information

The GfG Microtector III G888C is a compact personal multi-gas detector that continuously monitors the air around the wearer in diffusion mode. It holds up to five sensors and can display up to seven gas readings at once. GfG Instrumentation, Inc. in Ann Arbor, Michigan is GfG's US company.

The G888C is built for personal protection in hazardous or enclosed work areas. It samples by diffusion only. For measuring gas releases, searching for leaks or testing a confined space from outside, GfG recommends the pumped GfG Polytector III G999, which uses the same sensors, chargers, docking stations and radio system.

Key features

  • Five sensor positions: three electrochemical (toxic gases, O2 or H2), one catalytic bead for %LEL and one infrared position for CO2 and/or hydrocarbons.
  • Combination sensors (CO/H2S in one slot, or CO2 + %LEL in one IR sensor) allow up to seven readings from five positions.
  • Backlit full-graphic display that turns green, orange or red with alarm status, and flips 180° so it can be read while clipped on.
  • Three instantaneous alarm levels for O2 and combustibles; two for toxic gases plus STEL and TWA alarms.
  • 103 dB(A) horn (can be reduced to 90 dB(A)), vibration alarm and wraparound alarm LEDs.
  • Man-down alarm with two local warning levels.
  • Optional 915 MHz radio (North America) that sends readings, alarms and man-down status to a GfG-Link 10 or a Windows PC running G888/G999 Visual software.
  • Built-in LED flashlight. The same button can be set up as a manual (panic) alarm instead.
  • Alarm simulation for training, so readings can be pushed to the instrument without gas.
  • Data logger with a 1 second to 60 minute interval, read out through a docking station, test station or Smart Cap.
  • IP67 rubberized polycarbonate housing, up to 9.7 oz (275 g) depending on sensors.
GfG Microtector III G888C multi-gas detector with green display reading O2, H2, LEL, CO2, CO, H2S and NH3
GfG Microtector III G888C showing seven gas readings

Standard Sensor Options for the GfG G888

GasSensor typeMeasuring rangeResolutionExpected sensor lifeGfG USA part number
Oxygen (O2)Electrochemical0–25% vol0.1% vol5 years1990718
Combustibles (%LEL)Catalytic bead0–100% LEL or 0–5.00% vol CH40.5% LEL3 years1990851
Carbon monoxide (CO)Electrochemical0–500 ppm (1,000 or 2,000 ppm selectable)1 ppm3 years1990705
Hydrogen sulfide (H2S)Electrochemical0–100 ppm (50, 200 or 500 ppm selectable)0.1 ppm (0.2 ppm at 200, 0.5 ppm at 500)3 years1990700
CO + H2S combination (COSH)Electrochemical, one slot0–500 ppm CO / 0–100 ppm H2S1 ppm CO / 0.2 ppm H2S3 yearsConfirm US PN with GfG
Carbon dioxide (CO2)Infrared0–5% vol0.01–0.05% vol6 yearsConfirm US PN with GfG
Hydrocarbons (%LEL)Infrared0–100% LEL0.2–1.0% LEL6 yearsConfirm US PN with GfG
CO2 + hydrocarbonsInfrared combination0–5% vol CO2 and 0–100% LEL—6 yearsConfirm US PN with GfG

US part numbers for the O2, LEL, CO and H2S sensors are from GfG USA's G888 field operation manual (7004-888). GfG USA has not published US part numbers for the other sensors in its G888 documents. Ranges, resolutions and expected life are from GfG's G888 operating manual and are for clean air. Some sensors can only be used in specific electrochemical slots, so confirm the configuration with GfG when ordering. For reference only, the international manual lists O2 1990715/1990716/1990717 (2-, 3- and 5-year), standard LEL 1990850, COSH 1990710, CO2 1990900, hydrocarbon IR 1990905 and CO2 + hydrocarbon IR 1990920.

Other Electrochemical Sensors Available for the G888

GasMeasuring rangeResolution
Ammonia (NH3)0–300 ppm or 0–1,000 ppm1 ppm (0–300 ppm); 5 ppm (0–1,000 ppm)
Chlorine (Cl2)0–10 ppm (20 or 40 ppm selectable)0.05 ppm (0.1 ppm on 0–40 ppm)
Chlorine dioxide (ClO2)0–2 ppm0.01 ppm
Ethylene oxide (C2H4O)0–20 ppm (50 or 100 ppm selectable)0.1 ppm
Hydrogen (H2)0–2,000 ppm, 0–1.00% vol or 0–4.00% vol2 ppm / 0.01% vol
Hydrogen chloride (HCl)0–30 ppm (50 ppm selectable)0.2 ppm
Hydrogen cyanide (HCN)0–50 ppm (100 ppm selectable)0.1 ppm (0.2 ppm on 0–100 ppm)
Hydrogen fluoride (HF)0–10 ppm0.5 ppm
Nitric oxide (NO)0–100 ppm (50 or 200 ppm selectable)0.2–0.5 ppm
Nitrogen dioxide (NO2)0–30 ppm (50 or 100 ppm selectable)0.1 ppm
Phosgene (COCl2)0–2 ppm0.01 ppm
Phosphine (PH3)0–10 ppm (20 or 50 ppm selectable)0.05 ppm
Silane (SiH4)0–20 ppm (50 ppm selectable)0.1 ppm
Sulfur dioxide (SO2)0–10 ppm (20, 50 or 100 ppm selectable)0.05–0.1 ppm

GfG USA has not published US part numbers for these sensors in its G888/G999 documents, so confirm the part number with GfG when ordering. For reference only, GfG's international operating manual lists NH3 1990735/1990740, Cl2 1990725, ClO2 1990730, EtO 1990775, H2 1990785/1990790/1990795, HCl 1990755, HCN 1990760, HF 1990766, NO 1990745, NO2 1990750, COCl2 1990800, PH3 1990770, SiH4 1990780 and SO2 1990720.

Specifications

SpecificationGfG Microtector III G888C
SamplingDiffusion only
Sensor positions3 electrochemical, 1 catalytic, 1 infrared
DisplayBacklit full-graphic LCD; backlight green, orange or red by alarm status; rotates 180°
AlarmsO2 and combustibles: 3 instantaneous levels; toxics: 2 instantaneous plus STEL and TWA; battery alarm; man-down
Audible alarm103 dB(A), can be reduced to 90 dB(A)
BatteryRechargeable NiMH module, 2.6 V, 2,100 mAh
Run time (new battery, 68°F / 20°C, no radio)About 65 h (EC only); 23 h (EC + IR); 21 h (EC + catalytic with PowerSave); 13 h (EC + catalytic with PowerSave + IR)
Run time with radioAbout 26 h (EC); 15 h (EC + IR); 14 h (EC + catalytic with PowerSave); 10 h (EC + catalytic with PowerSave + IR)
Charge timeUp to 6 hours (drop-in charger or Smart Charger Cap); charge only in a non-hazardous area
Operating conditions-4 to +122°F (-20 to +50°C), 5–95% RH, 70–130 kPa (about 10–19 psia)
Storage-13 to +131°F (-25 to +55°C); 32 to 86°F (0 to 30°C) recommended
Dimensions2.7 x 3.9 x 1.5 in (68 x 100 x 39 mm), W x H x D
WeightUp to 9.7 oz (275 g), depending on sensors
Ingress protectionIP67
Radio (optional)915 MHz (US), about 985 ft (300 m) free-field range. EU versions use 868 MHz.
Data loggingAverage, peak or instantaneous values; 1 s to 60 min interval; ring memory
Hazardous location (North America)GfG's 2025 US data sheet lists c-CSA-us as pending for Class I, Division 1, Groups A, B, C and D, T3 and Class I, Zone 0 AEx ia IIC T3. Check the label on your instrument or confirm with GfG before use in a classified location.
Hazardous location (ATEX/IECEx)G888C: I M2 Ex ia db I Mb; II 2G Ex ia db IIC T4 Gb (-20°C ≤ Ta ≤ +50°C, i.e. -4 to +122°F); BVS 15 ATEX E 064 X; IECEx BVS 15.0056 X
Warranty (GfG USA)Limited lifetime on the instrument while in service; LEL, CO and H2S sensors 3 years; O2 sensor 5 years; consumable accessories 1 year

Bump Testing and Calibrating the GfG G888

  • GfG recommends a bump test before each day's use: expose the sensors to gas above the alarm set points and confirm the audible and visual alarms. Calibrate if readings are outside limits.
  • Calibrate before first use and then on a regular schedule. How often depends on how much the instrument is used and what the sensors are exposed to. Follow your company's program and any regulations that apply.
  • The instrument tracks bump, calibration and maintenance due dates and shows "Bump test/maintenance overdue" or "Recalibration needed" at start-up when one is past due.
  • Apply gas through the Smart Cap (1990210) at 0.5–0.6 LPM, the flow GfG specifies. Our 0.5 LPM Fixed Flow C10 Calibration Gas Regulator fits our 34 L, 58 L and 116 L aluminum and 103 L steel cylinders; choose the stainless steel version for reactive gases such as ammonia or chlorine.
  • Let the sensors warm up fully before calibrating, especially the catalytic sensor. About five minutes is usually enough.
  • The catalytic sensor is factory calibrated to 50% LEL methane. If you monitor a different flammable gas in the %LEL range, calibrate with that gas.

AutoCal procedure

  1. In clean air, turn the instrument on and let readings stabilize.
  2. Press the center and right buttons together to open the AutoCal menu, then press AIR. The instrument zeros the toxic and combustible sensors and sets O2 to 20.9%, then shows OK for each sensor that passes.
  3. Attach the Smart Cap and connect the regulator and tubing to the cylinder.
  4. Open AutoCal again, start the gas flow and press GAS. Calibration starts automatically after about 10 seconds and takes about 45 seconds. One cylinder of 4-gas mix calibrates the LEL, CO and H2S sensors at the same time.
  5. Confirm OK for each sensor, close the regulator and remove the Smart Cap before returning to service.
  6. If a sensor fails, repeat the step at least once. Check that the cylinder isn't empty and that the cylinder, tubing and cap are connected properly. GfG says a sensor that still fails must be replaced before the instrument is used.

Recommended calibration gas

SensorsCalibration gas (GfG specification)Matching Ideal Calibrations gas
O2, LEL (methane), CO, H2S50% LEL methane, 18% O2, 200 ppm CO, 20 ppm H2S, balance N2 (GfG 7802-030 / 7803-030 / 7804-030)Calibration Gas Blend 0481 (same composition)
Fresh-air adjustment (sets O2 to 20.9% and toxics/LEL to zero)Clean fresh air with no measurable contaminants. A cylinder of zero air is an alternative where clean air can't be assured (our suggestion, not a GfG requirement).Zero Air
CO2 (IR)4% vol CO2, balance N2 (GfG 7802-032). Because ambient air contains CO2, GfG's operating manual says to zero the CO2 sensor with CO2-free zero gas, such as synthetic air, 100% nitrogen or specially purified air.Carbon Dioxide (CO2) calibration gas (4% vol, N2 balance); Zero Nitrogen (N2) or Zero Air for zero
LEL on a gas other than methane50% LEL of the target gas, e.g. propane (GfG 7802-012)Propane (C3H8) calibration gas, Methane (CH4) calibration gas

For other toxic sensors, use single-gas cylinders at the concentrations GfG lists, for example Ammonia (NH3) calibration gas, Chlorine (Cl2) calibration gas, Nitrogen Dioxide (NO2) calibration gas, Sulfur Dioxide (SO2) calibration gas, Hydrogen Cyanide (HCN) calibration gas, Hydrogen (H2) calibration gas or Phosphine (PH3) calibration gas. The span value entered in the instrument must match the cylinder label. For a general walkthrough, see How to Calibrate Gas Detectors.

Catalytic LEL response to other gases

A catalytic sensor calibrated on methane does not read every flammable gas the same way. Heavier hydrocarbons such as propane, pentane or hexane typically read low on a methane calibration. GfG publishes response data in its manuals, but the figures differ between the US field manual and the international operating manual, which describe different catalytic sensor models. If you monitor a gas other than methane, calibrate with that gas or ask GfG for the response data for the sensor in your instrument.

Maintenance and Sensor Life

  • Clean the housing with a damp cloth and clear water only. Do not use solvents or detergents.
  • Keep the charging contacts clean, and charge or open the instrument only outside hazardous areas.
  • Protect the catalytic sensor from silicones, sulfur compounds, halogenated hydrocarbons and lead compounds. After a known exposure, check it with calibration gas. Do not test it with a butane lighter.
  • Above 110% LEL, the instrument shows over-range and switches off the catalytic sensor to protect it. Alarms stay on until reset in fresh air. Recalibrate, or replace the sensor if needed, after a high-concentration exposure.
  • Expected sensor life in clean air: CO and H2S about 3 years; O2 about 5 years (US replacement sensor 1990718); catalytic 3 years; infrared 6 years; most other toxic sensors 1–3 years.
  • Store spare sensors at 32–86°F (0–30°C). Use electrochemical spares within 6 months, and remove the battery if the instrument will be stored longer than 6 months.
  • Turning the instrument off resets TWA, STEL and peak values, so leave it on for the whole shift.

Accessories and Part Numbers

AccessoryGfG USA part numberNotes
Smart Charger Cap (charging, calibration, data transfer)1990210Fits over the diffusion inlets for calibration gas at 0.5–0.6 LPM
Smart Charger Cap with data interface cable1990212
Drop-in charger DIC 888/999, without plug-in power supply1990220Shared with the G999
Plug-in wall charger, 110 V AC (for the drop-in charger)4001-650
Vehicle charger4001-650VGfG Vehicle Charger (4001-650V)
USB data cable, drop-in charger to PC1990229
USB data cable, Smart Charger Cap to PC, 6 ft1450600
Data logger kit (software, cable, Smart Cap)1990205
Protective case with display cover and carrying straps (G888)1990297
RF transceiver USB dongle, 915 MHz, with G888/G999 Visual software (Windows)1990202For radio-equipped instruments
DS400 docking station (1 gas input)1450-406Shared with the G999
DS404 docking station (4 gas inputs)1450-407Shared with the G999

Related GfG Gas Detectors

  • GfG Polytector III G999: five-sensor monitor with a built-in pump for confined-space entry testing.
  • GfG G222E (Micro 5): 1- or 2-gas electrochemical detector.
  • G450 and G460: GfG Microtector II detectors. GfG describes the G450 as a 4-gas detector and the G460 as a multi-gas detector for up to 7 gases.
  • All GfG Instrumentation gas detector guides.

Reference documents

Information on this page comes from these GfG documents. Where they disagree, we use the newest US document and note the difference.

​