Product Overview
The RKI Instruments 33-0171RK is a replacement hydrophobic filter. RKI describes it as an AcroPak 300 with PTFE membrane, 0.2 µm. The hydrophobic PTFE membrane passes gas but blocks liquid water.
In the RKI Digester Gas Monitor it stops any water that gets past the water trap from entering the flow system.
Used On
- RKI Digester Gas Monitor (hydrophobic filter, listed in the operator’s manual parts list)
- RKI EAGLE (listed in RKI’s EAGLE replacement parts)
Which Part Number?
- 33-0171RK (this part): the replacement filter.
- 33-0171RK-01: the same filter installed internally (RKI price list).
- 33-0171RK-02: AcroPak 300 filter added to an EAGLE, field installed (RKI price list).
Specifications
Membrane and size data are the filter maker’s published values for AcroPak 300 capsules with 0.2 µm PTFE membrane.
| Specification | Value |
|---|---|
| Filter | AcroPak 300 capsule |
| Membrane | PTFE, 0.2 µm, hydrophobic, on a polypropylene support |
| Housing | Polypropylene |
| Effective filtration area | 280 cm² |
| Connections | Stepped hose barbs, 1/4" to 1/2" (6.4 to 12.7 mm) |
| Length | 4.1" (10.5 cm) with hose barbs |
| Maximum operating pressure | 60 psi (4.1 bar) at ambient temperature |
| Maximum operating temperature | 140°F (60°C) at 30 psi |
Replacing It in the Digester Gas Monitor
- Turn off the monitor and its incoming power, open the door, and turn the flow regulator knob fully counterclockwise.
- Pull the filter out of its metal clamp (just left of the main circuit board), then release the push fittings at the bottom and top.
- The replacement filter has rigid tubing at each end. Push the ends into the push fittings with the arrow pointing up.
- Set the filter back in the clamp, restore power, and set the sensor flow to 3 SCFH.
A dirty or clogged hydrophobic filter is one of the causes RKI lists for low Digester flow. Check the water trap first, then this filter.
Other Digester filters: oil mist filter 33-0164RK, particle filter 33-0167RK and water trap element 33-0413RK-02.
References
- RKI Instruments 2026 Price List (02/17/2026)
- RKI Digester Gas Monitor Operator’s Manual 71-0266RK (Rev. S, released 1/22/24)
- Pall / Cytiva data sheet: AcroPak 300 Capsule With PTFE Membrane (part numbers 12082 and 12085)
(P/N: 33-0171RK)