Product Overview
The RKI Instruments 33-0167RK is an inline particle filter marked CF-8369. RKI’s SDM manuals call it the “particle air filter replacement, CF-8369.” It keeps dust and particulates out of the sample or air path of the stations and detectors below.
Used On
- RKI SDM-03 calibration station (spare parts list)
- RKI SDM-04 calibration station
- RKI SDM-Force calibration station
- RKI SDM-2012 calibration station
- RKI 35-3001 and 35-3010 series sample-draw detectors (listed with the replacement parts; the 35-3001-01-01 manual lists it as the particle filter)
- RKI Digester Gas Monitor (particle filter, bottom right corner of the housing)
When to Replace
- Replace it when it is visibly dirty, or when low flow is traced to the filter.
- On the Digester Gas Monitor, RKI calls for a monthly visual check of all the sample filters.
Replacing It in the Digester Gas Monitor
- Turn off the monitor and its incoming power, open the door, and turn the flow regulator knob fully counterclockwise.
- Use pliers to remove the clamps at each end, then pull off the tubing. Replace damaged tubing with 3/16" ID x 5/16" OD tubing.
- Install the new filter in the same orientation. The arrow should point up, away from the Fresh Air Inlet fitting.
- Reinstall the clamps, restore power, and set the sensor flow to 3 SCFH.
The photo shows the filter installed with tubing and clamps. This listing is for the filter.
Other Digester filters: oil mist filter 33-0164RK, hydrophobic filter 33-0171RK and water trap element 33-0413RK-02.
References
- RKI Instruments 2026 Price List (02/17/2026)
- RKI Digester Gas Monitor Operator’s Manual 71-0266RK (Rev. S, released 1/22/24)
- RKI SDM-03 Operator’s Manual (stand-alone) and RKI SDM-Force Operator’s Manual, spare parts lists
- RKI 35-3001-01-01 Operator’s Manual, parts list
(P/N: 33-0167RK)