RKI SDM-3R Stand Alone Calibration Station with 58L calibration kit

Price: $1,875.00
SKU:
81-SDM3R103
Model:
GX-3R
mibo-boc:
3
Adding to cart… The item has been added

Product Overview

The RKI 81-SDM3R103 is an SDM-3R docking and calibration station for the GX-3R and GX-3R Pro portable gas monitors, supplied as a complete kit with a demand flow regulator and a 58-liter 4-gas calibration cylinder. The station automates charging, bump testing, and calibration, and operates as a stand-alone unit with no PC required. All records are stored automatically for later upload to a PC via the included USB flash drive.

Key Features

  • Stand-alone operation: Charges, bump tests, and calibrates without a PC connection, and turns the gas detector on and off automatically.
  • Automatic record keeping: Saves bump test, calibration, and alarm check records to a USB flash drive, computer, or network location.
  • PC-controlled option: Connected by USB cable, RKI's SDM Docking Station controller software can calibrate, bump test, alarm check, charge, and archive data for up to ten instruments at once, configure GX-3R (Pro) parameters, and download each instrument's datalogging history.
  • Single-cylinder configuration: This version is fitted with one solenoid for one calibration gas cylinder input. The SDM-3R has five ports: three calibration gas inputs, one fresh air inlet, and one exhaust. RKI offers two- and three-solenoid versions (81-SDM3R2xx and 81-SDM3R3xx) for applications that need more than one cylinder.
  • Modular design: Up to ten SDM-3R modules snap together and can share one set of calibration equipment, reducing the number of regulators and cylinders required.

What's Included (81-SDM3R103)

  • SDM-3R calibration station with 1 solenoid
  • Demand flow regulator
  • 58-liter aluminum 4-gas calibration cylinder (H₂S/CO/CH₄/O₂)
  • Cylinder holder
  • AC adapter
  • USB flash drive and USB cable (Type A to Type B)
  • Inlet air filter, 10 ft exhaust tube, and 3 ft gas fitting tube
  • PC software for viewing and archiving bump test, calibration, and event data

SDM-3R Kit Comparison

The 81-SDM3R103 and 81-SDM3R104 are the same single-solenoid station with demand flow regulator, cylinder holder, and 4-gas cylinder. The 81-SDM3R103 includes the larger 58-liter cylinder; the 81-SDM3R104 includes a 34-liter cylinder. The 81-SDM3R101 is the station without regulator or gas, and the 81-SDM3R102 adds the demand flow regulator without gas. The replacement 58-liter 4-gas cylinder for the GX-3R is RKI P/N 81-0154RK-02.

Specifications

  • Applicable instruments: GX-3R and GX-3R Pro
  • Input power: AC adapter, 100–240 VAC, 47–63 Hz, 0.4 A input; 5.99 VDC, 2 A output
  • Environment: indoor use only, 0–40 °C (32–104 °F), below 95% RH, non-condensing
  • Ports: 3 calibration gas, 1 fresh air, 1 exhaust
  • Stored records (stand-alone): 200 calibrations, 214 bump tests, or 469 alarm checks; unlimited when PC-controlled
  • PC software: Windows 7, 8, and 10 compatible
  • Warranty: 1 year

Docking-station maintenance keeps GX-3R and GX-3R Pro instruments on a consistent bump test and calibration schedule and produces a documented record of each test for compliance audits.

Compatible Gas Detectors

  • RKI GX-3R
  • RKI GX-3R Pro

(P/N: 81-SDM3R103)

Reviews

​