Product Overview
Replacement hydrogen cyanide (HCN) electrochemical sensor for the RAE Systems QRAE 3 (part 032-0205-000, 3R format). Fit it when the HCN channel fails calibration or reaches end of life.
QRAE 3 HCN channel (Honeywell)
- Instrument range: 0 to 50 ppm HCN, 0.2 ppm resolution
- Honeywell lists a 1-year warranty on the QRAE 3 HCN and SO2 sensors
- Honeywell’s sensor technical note lists 10 ppm HCN in nitrogen as the calibration gas for its HCN sensors
Things to know
- HCN sensors have strong cross-sensitivity to H2S and SO2. Honeywell’s note for QRAE 3-type HCN sensors flags a built-in filter to remove H2S; do not rely on the HCN reading where H2S may be present
- HCN is acutely toxic: calibrate only with a certified low-concentration cylinder and vent the exhaust safely
- Calibrate the QRAE 3 after replacing a sensor