RAE Systems Hydrogen Sulfide (H2S) Extended-Range Sensor, 0-1,000 ppm

Price: $330.00
SKU:
C03-0904-000
Model:
ToxiRAE Pro
Model:
MultiRAE Lite
Model:
AreaRAE Plus
Model:
AreaRAE Pro
Model:
BW RigRat
mibo-boc:
1
Adding to cart… The item has been added

Product Overview

The Honeywell RAE Systems C03-0904-000 is a 4R+ electrochemical hydrogen sulfide (H2S) extended-range (high-range) sensor. It has a measuring range of 0 to 1,000 ppm at 1 ppm resolution. It is a plug-in replacement sensor for the instruments listed below.

Check the instrument before ordering. This is the extended-range H2S sensor, not the standard-range H2S sensor C03-0907-001. Honeywell's MultiRAE replacement-sensor list restricts C03-0904-000 to MultiRAE diffusion models.

Compatible Instruments

Honeywell's January 2025 Sensors Table does not mark the RigRat column for this PN, but the same price list includes C03-0904-000 under BW RigRat spare sensors.

Not Listed For

Those MultiRAE instruments use C03-0907-001 for hydrogen sulfide.

The March 2018 TN-114 also named legacy AreaRAE, "MultiRAE Family," and QRAE Plus. Current Honeywell compatibility is the list above.

Difference from C03-0907-001

  • C03-0904-000 (this sensor): extended range, 0 to 1,000 ppm, 1 ppm resolution. Calibrate with 25 ppm H2S in nitrogen. On MultiRAE instruments, listed for diffusion models only.
  • C03-0907-001: standard range, 0 to 100 ppm, 0.1 ppm resolution. Calibrate with 10 ppm H2S in nitrogen. Used on pumped MultiRAE Lite, MultiRAE, and MultiRAE Pro.

Specifications

Specification Value
Sensor typeElectrochemical (4R+)
Gas detectedHydrogen sulfide (H2S), extended range
Range0 to 1,000 ppm
Resolution1 ppm
Bias / equilibrationNo bias; allow 10 minutes after installation
Temperature range-4°F to 122°F (-20°C to 50°C)
Pressure rangeAtmospheric ±10%
Operating humidity15 to 90% RH, non-condensing
Drift<2% signal/month
Storage life6 months in a sealed container
Storage temperature32°F to 68°F (0°C to 20°C)
Operating life2 years in air
Warranty2 years from date of shipment
Calibration gas25 ppm H2S, balance nitrogen

Installation, Warm-up, and Calibration

  1. Unbiased 4R+ sensors may ship with a shorting pin across the electrodes. Remove the pin before installation.
  2. Install the sensor in a compatible instrument.
  3. Allow 10 minutes for the reading to stabilize.
  4. Calibrate the instrument. Always calibrate after installing a new sensor. Honeywell specifies 25 ppm hydrogen sulfide in nitrogen as the span gas (the standard-range C03-0907-001 uses 10 ppm H2S in nitrogen).

See 25 ppm hydrogen sulfide (H2S) calibration gas, balance nitrogen.

Cross-Sensitivities

Published TN-114 cross-sensitivities for this extended-range sensor are low:

Gas Test concentration Typical response
Carbon monoxide (CO)300 ppm0 ppm
Ethylene100 ppm0 ppm
Hydrogen (H2)1,000 ppm0 ppm
Nitrogen dioxide (NO2)5 ppm0 ppm
Sulfur dioxide (SO2)5 ppm0 ppm
Nitric oxide (NO)35 ppmup to 3 ppm

Cross-sensitivity can vary by sensor and over its life.

(P/N: C03-0904-000)