Product Overview
The Honeywell RAE Systems C03-0904-000 is a 4R+ electrochemical hydrogen sulfide (H2S) extended-range (high-range) sensor. It has a measuring range of 0 to 1,000 ppm at 1 ppm resolution. It is a plug-in replacement sensor for the instruments listed below.
Check the instrument before ordering. This is the extended-range H2S sensor, not the standard-range H2S sensor C03-0907-001. Honeywell's MultiRAE replacement-sensor list restricts C03-0904-000 to MultiRAE diffusion models.
Compatible Instruments
- Honeywell RAE Systems ToxiRAE Pro (PGM-1860 toxic-gas models)
- Honeywell RAE Systems MultiRAE Lite, diffusion only (PGM-6208D), and other MultiRAE diffusion models
- Honeywell RAE Systems AreaRAE Plus
- Honeywell RAE Systems AreaRAE Pro
- Honeywell BW RigRat
Honeywell's January 2025 Sensors Table does not mark the RigRat column for this PN, but the same price list includes C03-0904-000 under BW RigRat spare sensors.
Not Listed For
- Pumped MultiRAE Lite (PGM-6208)
- MultiRAE (PGM-6228)
- MultiRAE Pro (PGM-6248)
- MicroRAE
- QRAE 3
Those MultiRAE instruments use C03-0907-001 for hydrogen sulfide.
The March 2018 TN-114 also named legacy AreaRAE, "MultiRAE Family," and QRAE Plus. Current Honeywell compatibility is the list above.
Difference from C03-0907-001
- C03-0904-000 (this sensor): extended range, 0 to 1,000 ppm, 1 ppm resolution. Calibrate with 25 ppm H2S in nitrogen. On MultiRAE instruments, listed for diffusion models only.
- C03-0907-001: standard range, 0 to 100 ppm, 0.1 ppm resolution. Calibrate with 10 ppm H2S in nitrogen. Used on pumped MultiRAE Lite, MultiRAE, and MultiRAE Pro.
Specifications
| Specification | Value |
|---|---|
| Sensor type | Electrochemical (4R+) |
| Gas detected | Hydrogen sulfide (H2S), extended range |
| Range | 0 to 1,000 ppm |
| Resolution | 1 ppm |
| Bias / equilibration | No bias; allow 10 minutes after installation |
| Temperature range | -4°F to 122°F (-20°C to 50°C) |
| Pressure range | Atmospheric ±10% |
| Operating humidity | 15 to 90% RH, non-condensing |
| Drift | <2% signal/month |
| Storage life | 6 months in a sealed container |
| Storage temperature | 32°F to 68°F (0°C to 20°C) |
| Operating life | 2 years in air |
| Warranty | 2 years from date of shipment |
| Calibration gas | 25 ppm H2S, balance nitrogen |
Installation, Warm-up, and Calibration
- Unbiased 4R+ sensors may ship with a shorting pin across the electrodes. Remove the pin before installation.
- Install the sensor in a compatible instrument.
- Allow 10 minutes for the reading to stabilize.
- Calibrate the instrument. Always calibrate after installing a new sensor. Honeywell specifies 25 ppm hydrogen sulfide in nitrogen as the span gas (the standard-range C03-0907-001 uses 10 ppm H2S in nitrogen).
See 25 ppm hydrogen sulfide (H2S) calibration gas, balance nitrogen.
Cross-Sensitivities
Published TN-114 cross-sensitivities for this extended-range sensor are low:
| Gas | Test concentration | Typical response |
|---|---|---|
| Carbon monoxide (CO) | 300 ppm | 0 ppm |
| Ethylene | 100 ppm | 0 ppm |
| Hydrogen (H2) | 1,000 ppm | 0 ppm |
| Nitrogen dioxide (NO2) | 5 ppm | 0 ppm |
| Sulfur dioxide (SO2) | 5 ppm | 0 ppm |
| Nitric oxide (NO) | 35 ppm | up to 3 ppm |
Cross-sensitivity can vary by sensor and over its life.
(P/N: C03-0904-000)